alveoair agrees closely with an established smart spirometer.
Sixty healthy people did the same breathing test on alveoair and on NuvoAir, a smart spirometer already cleared in the US and Europe. The two gave almost the same results for every main measurement.
- Design
- Prospective, cross-sectional, tertiary care centre
- Participants
- 60 healthy adults and students, aged 15 to 45
- Comparator
- NuvoAir smart spirometer
- Measures
- FEV1, FVC, PEF, FEV1/FVC
- Pearson r: FEV1 0.988, FVC 0.986, PEF 0.951, FEV1/FVC 0.91 (all p < 0.0001)
- ICC: FEV1 0.966, FVC 0.952, PEF 0.947, FEV1/FVC 0.91
- Bland-Altman: 95% to 98% of differences within the limits of agreement
- No statistically significant difference in means between devices
Citation and limitations
Shinde PS, Kulkarni RR, Kulkarni GV. Validation of portable Bluetooth enabled smart spirometer (Alveoair) for the measurement of various lung functions in healthy subjects. Natl J Physiol Pharm Pharmacol. 2021;11.
Limitations. Healthy participants only, at a single centre. The comparator is a portable device, not a laboratory spirometer. Patients with lung disease need further study.